Beam Test of FARICH Prototype with Digital Photon Counter Научная публикация

Общая информация Язык: Английский, Жанр: Статья (Full article),
Статус опубликования: Опубликована, Оригинальность: Оригинальная
Конференция 13th Vienna Conference on Instrumentation (VCI2013)
11-15 февр. 2013 , Vienna
Журнал Nuclear Instruments and Methods in Physics Research Section A
ISSN: 0168-9002
Вых. Данные Год: 2013, Том: 732, Страницы: 352-356 Страниц : 5 DOI: 10.1016/j.nima.2013.07.068
Ключевые слова Aerogel, Particle identification, Ring imaging Cherenkov counter
Авторы Barnyakov A.Yu. 1 , Barnyakov M.Yu. 1,3 , Bobrovnikov V.S. 1 , Buzykaev A.R. 1 , Danilyuk A.F. 4 , Degenhardt C. 6 , Dorscheid R. 6 , Finogeev D.A. 5 , Frach T. 6 , Gulevich V.V. 1 , Karavicheva T.L. 5 , Kononov S.A. 1,2 , Kravchenko E.A. 1,2 , Kurepin A.B. 5 , Kuyanov I.A. 1 , Muelhens O. 6 , Onuchin A.P. 1,3 , Ovtin I.V. 1,3 , Razin V.I. 5 , Reshetin A.I. 5 , Schulze R. 6 , Talyshev A.A. 1,2 , Usenko E.A. 5 , Zwaans B. 6
Организации
1 Budker Institute of Nuclear Physics SB RAS, Novosibirsk, Russia
2 Novosibirsk State University, Russia
3 Novosibirsk State Technical University, Russia
4 Boreskov Institute of Catalysis SB RAS, Novosibirsk, Russia
5 Institute of Nuclear Research RAS, Moscow, Russia
6 Philips Digital Photon Counting, Aachen, Germany

Информация о финансировании (2)

1 Сибирское отделение Российской академии наук 103
2 Министерство образования и науки Российской Федерации

Реферат: In June 2012 we tested a Focusing Aerogel RICH detector prototype based on Digital Photon Counters (DPC) by Philips at the CERN PS T10 beam line with a particle momentum up to 6 GeV/c. The prototype comprises a 20×20 cm2 photon detector with 48×48 DPC pixels. In order to reduce the dark count rate, the photon detector was cooled to −40 °C in addition to disabling individual DPC's microcells. A four layer focusing aerogel radiator with 200 mm focal distance was studied. We obtained a Cherenkov angle resolution of 3.61±0.04 mrad. The mean number of photoelectrons in a ring is 12. Directly measured π/K separation at 6 GeV/c momentum is 3.5σ, μ/π separation is 5.3σ at 1 GeV/c. A comparison with a Monte Carlo simulation is presented as well.
Библиографическая ссылка: Barnyakov A.Y. , Barnyakov M.Y. , Bobrovnikov V.S. , Buzykaev A.R. , Danilyuk A.F. , Degenhardt C. , Dorscheid R. , Finogeev D.A. , Frach T. , Gulevich V.V. , Karavicheva T.L. , Kononov S.A. , Kravchenko E.A. , Kurepin A.B. , Kuyanov I.A. , Muelhens O. , Onuchin A.P. , Ovtin I.V. , Razin V.I. , Reshetin A.I. , Schulze R. , Talyshev A.A. , Usenko E.A. , Zwaans B.
Beam Test of FARICH Prototype with Digital Photon Counter
Nuclear Instruments and Methods in Physics Research Section A. 2013. V.732. P.352-356. DOI: 10.1016/j.nima.2013.07.068 WOS Scopus РИНЦ