A Compact Device for Loading Diamond Anvil Cells with Low-Boiling Pressure-Transmitting Media Научная публикация

Журнал Journal of Applied Crystallography
ISSN: 0021-8898 , E-ISSN: 1600-5767
Вых. Данные Год: 2013, Том: 46, Номер: 1, Страницы: 267-269 Страниц : 3 DOI: 10.1107/S0021889812046912
Ключевые слова diamond anvil cell loading, high-pressure crystallography, high-pressure instrumentation, pressure-transmitting media
Авторы Zakharov B.A. 1,2 , Achkasov A.F. 2
Организации
1 Institute of Solid State Chemistry and Mechanochemistry, SB RAS, Kutateladze Street 18, 630128, Novosibirsk, Russian Federation
2 Novosibirsk State University, Pirogova Street 2, 630090, Novosibirsk, Russian Federation

Реферат: A new compact laboratory device for loading diamond anvil cells with low-boiling pressure-transmitting media is described. This device allows, in particular, the easy loading of diamond anvil cells with pentane-isopentane (1:1) mixture at environmental temperatures up to 303 K and relative humidity at least up to 75-80%. The relatively simple design, the easy availability of materials used for the device and the application of liquid nitrogen as cooling agent allow the manufacture of such a device in a short time at almost any laboratory. © 2013 International Union of Crystallography.
Библиографическая ссылка: Zakharov B.A. , Achkasov A.F.
A Compact Device for Loading Diamond Anvil Cells with Low-Boiling Pressure-Transmitting Media
Journal of Applied Crystallography. 2013. V.46. N1. P.267-269. DOI: 10.1107/S0021889812046912 WOS Scopus РИНЦ CAPlusCA OpenAlex
Даты:
Поступила в редакцию: 5 окт. 2012 г.
Принята к публикации: 14 нояб. 2012 г.
Опубликована online: 21 дек. 2012 г.
Опубликована в печати: 1 февр. 2013 г.
Идентификаторы БД:
Web of science: WOS:000313658700034
Scopus: 2-s2.0-84872716506
РИНЦ: 20420526
Chemical Abstracts: 2013:111769
Chemical Abstracts (print): 158:561731
OpenAlex: W1965913748
Цитирование в БД:
БД Цитирований
Web of science 14
Scopus 14
OpenAlex 11
Альметрики: