High Density Aerogel for ASHIPH SND-Test Results Научная публикация

Конференция 10th International Conference on Instrumentation for Colliding Beam Physics
28 февр. - 5 мар. 2008 , Novosibirsk
Журнал Nuclear Instruments and Methods in Physics Research Section A
ISSN: 0168-9002
Вых. Данные Год: 2009, Том: 598, Номер: 1, Страницы: 163-165 Страниц : 3 DOI: 10.1016/j.nima.2008.08.018
Ключевые слова Aerogel, Cosmic tests, Particle identification, Time resolution
Авторы Barnyakov A.Yu. 1 , Barnyakov M.Yu. 1 , Beloborodov K.I. 1,3 , Bobrovnikov V.S. 1 , Buzykaev A.R. 1 , Danilyuk A.F. 2 , Golubev V.B. 1,3 , Kirillov V.L. 2 , Kononov S.A. 1 , Kravchenko E.A. 1,3 , Onuchin A.P. 1 , Martin K.A. 1,4 , Serednyakov S.I. 1,3 , Vesenev V.M. 1
Организации
1 Budker Institute of Nuclear Physics, Novosibirsk, Russia
2 Boreskov Institute of Catalysis, Novosibirsk, Russia
3 Novosibirsk State University, Novosibirsk, Russia
4 Novosibirsk State Technical University, Novosibirsk, Russia

Информация о финансировании (8)

1 Российский фонд фундаментальных исследований 06-02-16192
2 Российский фонд фундаментальных исследований 06-02-16273
3 Российский фонд фундаментальных исследований 06-02-16294
4 Российский фонд фундаментальных исследований 07-02-00104
5 Российский фонд фундаментальных исследований 08-02-00328
6 Российский фонд фундаментальных исследований 08-02-00634
7 Российский фонд фундаментальных исследований 08-02-00660
8 Совет по грантам Президента Российской Федерации НШ-5655.2008.2

Реферат: The paper is devoted to the cosmic tests of particle identification system for the SND detector at VEPP-2000. The goal of the tests is to measure main parameters of the system: average signal for the relativistic particles, nonuniformity of the light collection, and time resolution. The measurements of the average signal were done in five points of the counter. The obtained value of the signal is about 10 photoelectrons. The time resolution is about 2 ns. The time dependence of the average amplitude was measured also. It was shown that average amplitude of the signal decreases with characteristic time of about two years. The cause of the signal decrease is discussed.
Библиографическая ссылка: Barnyakov A.Y. , Barnyakov M.Y. , Beloborodov K.I. , Bobrovnikov V.S. , Buzykaev A.R. , Danilyuk A.F. , Golubev V.B. , Kirillov V.L. , Kononov S.A. , Kravchenko E.A. , Onuchin A.P. , Martin K.A. , Serednyakov S.I. , Vesenev V.M.
High Density Aerogel for ASHIPH SND-Test Results
Nuclear Instruments and Methods in Physics Research Section A. 2009. V.598. N1. P.163-165. DOI: 10.1016/j.nima.2008.08.018 WOS Scopus РИНЦ CAPlus OpenAlex
Даты:
Опубликована online: 17 авг. 2008 г.
Опубликована в печати: 1 янв. 2009 г.
Идентификаторы БД:
Web of science: WOS:000262662500040
Scopus: 2-s2.0-57649099863
РИНЦ: 13606638
Chemical Abstracts: 2008:1493626
OpenAlex: W1984964769
Цитирование в БД:
БД Цитирований
Web of science 37
Scopus 45
РИНЦ 48
OpenAlex 39
Альметрики: