Calorimeter for Measurement of Heat of Chemisorption of Gases at 25-400 C-1 Full article
Journal |
Kinetics and Catalysis
ISSN: 0023-1584 , E-ISSN: 1608-3210 |
||
---|---|---|---|
Output data | Year: 1968, Volume: 9, Number: 6-2, Pages: 1129-1134 Pages count : 6 | ||
Tags | Absorption; Calorimeters | ||
Authors |
|
||
Affiliations |
|
Abstract:
Article describes differential double calorimeter with compensation of measured heat effect. Calorimeter has two adjustable shells, and sensitivity of 1. 5. 10-//5 degree. Absolute error of measurements is 6. 10-//4 cal/hr. The device is applied to study of dependence of heat of reaction of CO with NiO surface on amount of reacted carbon monoxide.