Sciact
  • EN
  • RU

FARICH Detector Beam Test Results Full article

Conference The International Conference "Instrumentation for Colliding Beam Physics"
24-28 Feb 2020 , Novosibirsk
Journal Journal of Instrumentation
ISSN: 1748-0221
Output data Year: 2020, Volume: 15, Number: 10, Article number : C10014, Pages count : 10 DOI: 10.1088/1748-0221/15/10/c10014
Tags Cherenkov detectors; Materials for solid-state detectors; Particle identification methods; Photon detectors for UV, visible and IR photons (solid-state)
Authors Barnyakov A.Yu. 1,3,4 , Barnyakov M.Yu. 1 , Boborovnikov V.S. 1,3 , Buzykaev A.R. 1,3 , Bykov A.V. 1 , Danilyuk A.F. 2,3 , Kasyanenko P.V. 1,3 , Katcin A.A. 1,3 , Kononov S.A. 1,3 , Kravchenko E.A. 1,3 , Kuyanov I.A. 1,3 , Onuchin A.P. 1,3 , Ovtin I.V. 1,3 , Podgornov N.A. 1,3 , Porosev V.V. 1,3 , Shalygin A.S. 2 , Shekhtman L.I. 1,3 , Talyshev A.A. 1,3
Affiliations
1 Budker Institute of Nuclear Physics, Acad. Lavrentiev Prospect 11, Novosibirsk 630090, Russia
2 Boreskov Institute of Catalysis, Acad. Lavrentiev Prospect 5, Novosibirsk 630090, Russia
3 Novosibirsk State University, St. Pirogova 2, Novosibirsk 630090, Russia
4 Novosibirsk State Technical University, Karl Marks Prospect 20, Novosibirsk 630073, Russia

Funding (2)

1 Russian Foundation for Basic Research 18-29-06022
2 Russian Foundation for Basic Research 18-32-00905 (АААА-А18-118041190030-0)

Abstract: The excellent particle identification (PID) system is needed for the successful execution of the broad experimental program at future Super C-τ Factory (SCTF) in Novosibirsk. The main requirements for the PID system are the following: good π/K-separation in whole operational momentum range and good µ/π-separation in the momentum range from 0.4 up to 1.2 GeV/c. The RICH detector based on focusing aerogel (FARICH) could provide good π/K-separation from 0.4 GeV/c and µ/π-separation in the momentum range from 0.4 up to 1.5 GeV/c. The method FARICH is described, beam test results are presented and the status of multilayer aerogel production is given
Cite: Barnyakov A.Y. , Barnyakov M.Y. , Boborovnikov V.S. , Buzykaev A.R. , Bykov A.V. , Danilyuk A.F. , Kasyanenko P.V. , Katcin A.A. , Kononov S.A. , Kravchenko E.A. , Kuyanov I.A. , Onuchin A.P. , Ovtin I.V. , Podgornov N.A. , Porosev V.V. , Shalygin A.S. , Shekhtman L.I. , Talyshev A.A.
FARICH Detector Beam Test Results
Journal of Instrumentation. 2020. V.15. N10. C10014 :1-10. DOI: 10.1088/1748-0221/15/10/c10014 WOS Scopus РИНЦ OpenAlex
Dates:
Submitted: May 14, 2020
Accepted: Jul 6, 2020
Published print: Oct 9, 2020
Published online: Oct 9, 2020
Identifiers:
Web of science: WOS:000577277600014
Scopus: 2-s2.0-85094875554
Elibrary: 45213789
OpenAlex: W3092542842
Citing:
DB Citing
Scopus 2
Elibrary 3
OpenAlex 2
Altmetrics: