The Use of High-Speed Multicapillary Column in ComprehensiveTwo-Dimensional Gas Chromatography with Flow Modulation Научная публикация

Журнал Journal of Chromatography A
ISSN: 0021-9673 , E-ISSN: 1873-3778
Вых. Данные Год: 2015, Том: 1426, Страницы: 183-190 Страниц : 8 DOI: 10.1016/j.chroma.2015.11.064
Ключевые слова Comprehensive two-dimensional chromatography, High-speed GC, Multicapillary chromatography, Multicapillary column
Авторы Patrushev Yuri V. 1,2 , Sidelnikov Vladimir N. 1
Организации
1 Boreskov Institute of Catalysis, pr. Lavrentieva 5, Novosibirsk 630090, Russia
2 Novosibirsk State University, Pirogova Str. 2, Novosibirsk 630090, Russia

Реферат: The possibility to use multicapillary gas-chromatography column (MCC) as the second-dimension column in comprehensive two-dimensional gas chromatography (GC × GC) has been demonstrated. Dependences of the efficiency on the carrier gas volumetric flow rate were studied for capillary columns (CC) HP-5 and ZB-WAX of length 5 m and multicapillary columns of length 0.25 m with stationary phases OV-73 and Carbowax 20 M. For standard capillary columns, the maximum efficiency is reached at a flow rate below 2 mL/min. In the GC × GC mode, a capillary column is operated at a flow rate strongly exceeding the optimal value. Therewith, the efficiency of standard capillary column decreases several-fold, to 1570 theoretical plates for HP-5 and 1000 theoretical plates for ZB-WAX. In the case of multicapillary columns that are used in this paper, the maximum efficiency is reached at the carrier gas velocities above 60 mL/min. This allows MCC to be used as the second-dimension column in GC × GC at the optimal flow rate, thus reaching a maximum efficiency of 5900 theoretical plates for MCC with the OV-73 stationary phase and 2800 theoretical plates with Carbowax 20 M. The GC × GC separation of aliphatic and aromatic compounds was used to demonstrate that high efficiency of MCC enhances peak resolution in the second dimension as compared to standard capillary column. MCC was shown to be superior in separation of the components with close retention times and substantially differing concentrations
Библиографическая ссылка: Patrushev Y.V. , Sidelnikov V.N.
The Use of High-Speed Multicapillary Column in ComprehensiveTwo-Dimensional Gas Chromatography with Flow Modulation
Journal of Chromatography A. 2015. V.1426. P.183-190. DOI: 10.1016/j.chroma.2015.11.064 WOS Scopus РИНЦ CAPlus OpenAlex
Даты:
Поступила в редакцию: 1 окт. 2015 г.
Принята к публикации: 17 нояб. 2015 г.
Опубликована online: 22 нояб. 2015 г.
Опубликована в печати: 1 дек. 2015 г.
Идентификаторы БД:
Web of science: WOS:000367276100020
Scopus: 2-s2.0-84955236860
РИНЦ: 26823268
Chemical Abstracts: 2015:1916501
OpenAlex: W2173790376
Цитирование в БД:
БД Цитирований
Web of science 11
Scopus 11
РИНЦ 10
OpenAlex 11